Hanna Instruments HI 4421 User Manual

Page 23

background image

23

Calibration

Using DO standards

The meter can be calibrated in a single or multi-points (up to two points), using 2 DO standards (0.0%, 100%)
or using the custom standards.
The following options are available for calibration:

Standard Recognition

The user can choose between

Automatic

recognition (from 2 Hanna standards available) or

User Standard

.

• Use

or

to highlight the

Standard Recognition

option.

• Use

or

to choose from

Automatic

or

User

Standard

option

.

• Press

to confirm your selection or press

to

cancel operation.

Calibration Reminder

This option allows the user to set the calibration reminder as

Daily

,

Periodic

or

Disabled

.

To set the calibration reminder:
• Use

or

to highlight the

Calibration re-

minder

option.

• Press

to confirm your selection and then use

or

to choose the desired option

.

• Press

to confirm your selection or press

to

cancel operation.