Result screen

The ion concentration is displayed in g/L (ppt) K

+

. If the

concentration is out of range, the instrument displays the upper

range limit blinking.

Press LOG to save the measured concentration. A message

indicating the free log space will be displayed for a few seconds.

If the free log space is under 10% this message will be

permanently displayed.

Press Start to begin a new measurement cycle.
Wrong slope screen

This screen appears if the computed slope isn’t in the

acceptable range. The acceptable range is 30% to 140%

default slope (53mV/decade).

Press Start to begin a new measurement cycle.

Add standard screen

Add the corresponding standard. Press READ to start the next

step analysis if “Reading confirm” option is enabled in the

Setup menu. If “Reading confirm” option is disabled, the

instrument will start the next step analysis.

Wrong standard error screen

This screen is displayed if step 2 and 3 of the current reading is

out of ±2000mV range or if the difference between the stable

readings from the current and the previous step is too small

(±3mV) or, if in auto confirm mode, the standard addition is

not detected.

Wrong input error screen

This screen is displayed if the current reading is out of

±2000mV range.

Pause screen

Press Resume to continue the measurement process.

Press Stir to start/stop the stirrer.

Sample reading confirm screen (only if this feature is enabled

in setup)

Press CFM to memorize the reading.