beautypg.com

Measurement procedure, Tips for an accurate measurement – Hanna Instruments HI 84181 User Manual

Page 14

background image

14

MEASUREMENT PROCEDURE

• Fill the beaker up to the 50 mL mark with sample.

Use the 2000 µL automatic pipette to add 2 mL of

HI 84181-0 ISA, place the stir bar into the beaker and

then place the beaker in the appropriate place on the

instrument top.

• Place the electrode holder on the top of the beaker

and secure it by turning clockwise.

• Immerse HI 61014 potassium electrode, HI 5315

reference and the HI 7662-T temperature probe

approximately 2 cm (0.8“) into the sample to be

tested paying attention to not touch the stir bar, and

press Start.

• Wait until the reading is stable, CFM key displayed, if

the “Reading confirm” option is selected, and then

press CFM. If the “Reading confirm” option is

disabled the instrument automatically memorizes the

reading when this is stable and waits to add standard.

Until the standard addition is detected the instrument

will beep if beeper is enabled in Setup.

TIPS FOR AN ACCURATE MEASUREMENT

The instructions listed bellow should be carefully followed during analysis to ensure best accuracy.
• Rinse the electrode pair with distilled or deionized water between samples, and blot dry with

a lab wipe or other soft disponible absorbent toweling.

• Select the adequate stability criteria.
• Analyse the wine sample at room temperature.