PRINCIPLE OF OPERATION

The determination of potassium concentration is made according to double standard addition method.
This technique enables an unknown concentration and the electrode’s slope factor to be determined
simultaneously.
The Automatic Temperature compensation is available using HI 7662-T temperature probe. By
setting the electrode isopotential point and enabling temperature compensation option in Setup
menu, the obtained concentration is temperature compensated.
The accuracy of the obtained result depends on the stability of the mV reading in every step of the
measurement.
Three selectable stability criteria can be selected in Setup menu. Select Medium or Accurate
criteria in order to have accurate analysis.

Range

0.0 to 5.0 g/L (ppt) K

+

Resolution

0.1 g/L (ppt) K

+

Accuracy

±5% of reading (see note)

Sample volume

50 mL

Temperature Compensation

Automatic from 0 to 80 °C

Electrodes

HI 61014 Potassium half cell

HI 5315 Reference half cell

Temperature Probe

HI 7662-T (included)

LOG feature

50 samples

Stirring speed

500 rpm

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condensing

Power supply

220V/50Hz; 10VA

Dimensions

208 x 214 x 163 mm (8.2 x 8.4 x 6.4") (with beaker)

Weight

2200 g (77.6 oz.)

Note: Above 2.5 g/L the instrument provides information about the approximate potassium content.

REQUIRED REAGENTS
Code

Description

Quantity/Test

HI 84181-0

ISA

2 mL

HI 84181-1

Standard 1

2 mL

HI 84181-2

Standard 2

2 mL

SPECIFICATIONS