Procedure:
• Switch the instrument on after connecting the pH

electrode.

• Rinse and immerse the pH electrode in pH 7.01

buffer and shake briefly. Wait 1 or 2 minutes for
the reading to stabilize.

• Note the temperature of the buffer solution using

a ChecktempC or a glass thermometer (e.g.20°C).

• Press the °C key to select temperature setting. Ad-

just the Temperature °C knob until display shows
the noted temperature.

• Press the pH key to select pH measure-

ment.

• Wait 1-2 minutes and adjust the OFFSET

knob until display shows the pH value at
the noted temperature (see the pH versus
temperature chart at page 23).

• Rinse and immerse the pH electrode in

pH 4.01/pH 10.01 buffer and shake
briefly.

• Wait 1-2 minutes and adjust the SLOPE

knob until display shows the pH value at
the noted temperature (see the pH versus
temperature chart at page 23).

• The pH calibration is now complete.