beautypg.com

Hanna Instruments HI 8521 User Manual

Page 11

background image

11

off. Press the RESET button and restart the entire operation.
Power connection
Plug the 12VDC adapter (HI710005 or HI710006) into the socket on
the back of the meter (#5).

Note: Make sure the main line is protected by a fuse.

Note: HI8520 uses a rechargeable battery to retain the pH calibra-

tion data. It lasts approximately one month when fully charged.
The battery is automatically charged any time the meter is
plugged-in (either turned on or off).

SPECIFICATIONS

HI 8520

Range

pH

0.00 to 14.00

°C

0.0 to 100.0

Resolution

pH

0.01

°C

0.1

Accuracy

pH

±0.01

(@20°C / 68°F) °C

±0.4

Typical EMC

pH

±0.02

Deviation

°C

±0.4

Calibration

Automatic 2 points with auto-buffer

recogntion

Temperature

Automatic or Manual

Compensation

From 0 to 100°C (32 to 212°F)

Electrode

HI 1332B plastic body, combination,

double junction refillable

pH electrode (included)

Temperature Probe

HI 7669/2W temp. probe (included)

Input Impedance

10

12

ohm

Power

12VDC power

Environment

0 to 50°C (32 to 122°F); 95% RH

Dimensions

230x170x70 mm (9.1x6.7x2.7")

Shipping Weight

1.35 Kg (3 lb.)