Functional description & specifications of hi8521, The front panel of hi8521, The rear panel of hi8521 – Hanna Instruments HI 8521 User Manual

Page 12

background image

12

COMBINED

pH or ORP

ELECTRODES

REFERENCE
ELECTRODE RESET

TEMPERATURE
PROBE

1

2

3

4

5

DC

POWER

12V

to

20V

+

RANGE

OFF

Temperature

Buffer Values (pH)

°C

°F

4.01

7.01

10.01

0

10
20
30
40

32
50
68
86

104

4.01
4.00
4.00
4.02
4.04

7.13
7.07
7.03
7.00
6.98

10.32
10.18
10.06

9.96
9.88

CON

ERROR SYMBOLS

°C

CAL

ON

E1: pH out of range

E2: °C out of range

E3: mV out of range

E4: wrong pH 7

E5: wrong pH 4/10

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI8521

THE FRONT PANEL OF HI8521

ON / OFF

To switch instrument ON and OFF.

RANGE

To display pH / ORP / Temperature measurements.

CAL/CON

CAL to enter the calibration procedure and CON to
confirm calibration data.

/

For setting temperature for manual compensation.

THE REAR PANEL OF HI8521

1. BNC socket for pH or ORP electrode.
2. Reference Jack socket for reference electrode.
3. RESET button.
4. Temperature Probe socket (for HI7669/2W).
5. DC Power Socket (for HI710005 or HI710006).

Electrode and probes connections
For combination pH or ORP electrodes (with internal reference) plug
the electrode's BNC to the socket provided (#1).
For an electrode with a separate reference, connect the measuring
electrode's BNC to the BNC socket and the reference electrode's jack
to the socket provided (#2).
For temperature measurements and automatic temperature com-
pensation connect the temperature probe to the °C socket (#4).
RESET button
The RESET button (#3) is used when the instrument displays errone-
ous messages due to strong electrical interference or when the