instrument's power supply is disconnected before the meter is switched
off. Press the RESET button and restart the entire operation.
Power connection
Plug the 12VDC adapter (HI710005 or HI710006) into the socket on
the back of the meter (#5).
Note: Make sure your main line is protected by a fuse.
Note: HI8521 uses a rechargeable battery to retain the pH calibra-
tion data. It lasts approximately one month when fully charged.
The battery is automatically charged any time the meter is
plugged-in (either turned on or off).
SPECIFICATIONS
HI 8521
Range
pH
0.00 to 14.00
ISE
0.0 to ±399.9mV
ORP
0 to ±1999mV
°C
0.0 to 100.0
Resolution
pH
0.01
ISE
0.1mV
ORP
1mV
°C
0.1
Accuracy
pH
±0.01
(@20°C / 68°F) ISE
±0.2mV
ORP
±0.1mV
°C
±0.4
Typical EMC
pH
±0.05
Deviation ISE/ORP
±4mV
°C
±0.1
Calibration
Automatic 2 points with auto-buffer
recogntion
Temperature
Automatic or Manual
Compensation
From 0 to 100°C (32 to 212°F)
Electrode
HI 1332B plastic body, combination,
double junction refillable
pH electrode (included)
Temperature Probe
HI 7669/2W temp. probe (included)
Input Impedance
10
12
ohm
Power
12VDC power
Environment
0 to 50°C (32 to 122°F); 95% RH
Dimensions
230x170x70 mm (9.1x6.7x2.7")
Shipping Weight
1.4 Kg (3.1 lb.)