PRINCIPLE OF OPERATION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and
matter. When a light beam crosses a substance, some of the radiation may be absorbed by
atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length
through the matter and on the

physical

-chemical characteristics of the substance according to the

Lambert-Beer Law:

-log

I

/

I

o

=

ε

λ

c d

or

A

=

ε

λ

c d

Where:

-log

I

/

I

o

=

Absorbance (A)

I

o

=

intensity of incident light beam

PRECISION AND ACCURACY

Precision

is how closely repeated measurements

agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy

is defined as the nearness of a test

result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of the
parameter and a representative lot of reagent
(for each parameter) the following standard
deviations were obtained:

Parameter

Standard Solution

Standard Deviation

pH

7.0 pH

0.1 pH units

Free Chlorine

1.00 mg/L

0.03 mg/L

Total Chlorine

1.00 mg/L

0.03 mg/L

Cyanuric Acid

20 mg/L

1 mg/L