• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for measurement.

• Remove the cuvette and add one packet

of HI 93711-0 reagent. Replace the cap
and shake gently for 20 seconds (or
2 minutes in case of seawater analysis).

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for

three seconds.

The display will show

the countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.

• Alternatively, wait for 2 minutes and 30

seconds and just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• The instrument directly displays

concentration in mg/L of total chlorine
on the Liquid Crystal Display.

INTERFERENCES

Interference may be caused by:

Bromine, Iodine, Chlorine Dioxide, Ozone, Oxidized manganese and Chromium.

In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for approximately

1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO

3

or acidity greater than 150 mg/L

CaCO

3

, the color of the sample could disappear or develop only partially. To resolve this,

neutralize the sample with diluted HCl or NaOH.