Hanna Instruments HI 96104 User Manual

Page 20

background image

20

• Place the corresponding CAL CHECK

TM

Standard Cuvette B into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.

• Press CAL CHECK and the lamp, cuvette and

detector icons together with “CAL CHECK”
will appear on the display, depending on the
measurement phase.

• At the end of the measurement the display

will show the validation standard value.

The reading should be within specifications as reported in the CAL CHECK

TM

Standard

Certificate. If the value is found out of the specifications, please check that the cuvettes are free
of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications,
then recalibrate the instrument.