• Rel mV calibration data (if available), which contains:

• the calibration offset, with sign (7 chars)
• the calibration time, yymmddhhmmss (12 chars).

PAR

Requests the setup parameters setting.

The answer string contains:

• Instrument ID (4 chars)
• Calibration alarm time out (2 chars)
• SETUP information (2 chars): 8 bit hexadecimal

encoding.
• 0x01 - beep ON (else OFF)
• 0x04 - degrees Celsius (else degrees Fahrenheit)
• 0x08 - Offset calibration (else Point calibration)

• Auto-off/Light time (3 chars)
• Auto power off time (3 chars)
• The number of custom buffers (1 char)
• The custom buffer values, with sign and decimal

point, for each defined custom buffer (7 chars)

NSLx

Requests the number of logged samples (4 chars).

The command parameter (1 char):

• P - request for pH range
• M - request for mV and Rel mV ranges

LODPxxx Requests the xxx

th

pH record logged data.

LODMxxx Requests the xxx

th

mV/Rel mV record logged data.

LODPALL Requests all pH Log on demand.
LODMALL Requests all mV/Rel mV Log on demand.
The answer string for each record contains:

• The logged mode (2 chars):

• 01 - pH range (0.01 resolution)
• 02 - pH range (0.1 resolution)
• 03 - mV range
• 04 - Rel mV range

• Reading status (1 char): R, O, U
• Calculated reading, with sign and decimal point

(7 chars) - for pH and Rel mV range

• Temperature reading, with sign and two decimal

points (8 chars)

• mV reading status (1 char): R, O, U - only for

HI 98150 and HI 98160