• If “WRONG” tag and “OLd” message are displayed on

the secondary LCD, an inconsistency exists between new

and previous (old) calibration data. Clear old calibration

parameters and proceed with calibration from the current

calibration point. The instrument will keep all confirmed

values during current calibration.

• With one-point calibration there is no “Condition” and only

the frame is shown. Calibration time out is active.

• Each time a buffer is confirmed, the new calibration

parameters replace the old calibration parameters of

the corresponding buffer.
If current confirmed buffer has no

correspondence in the existing stored

calibration and this is not full, the

current buffer is added to the existing

stored calibration.
If the existing stored calibration is

full (five calibration points), after confirming the last

calibration point, the instrument will ask which buffer will be

replaced by current buffer.
Press the ARROW keys to select another buffer to be replaced.
Press CFM to confirm the buffer that will be replaced.
Press CAL to exit this function. In this case, the buffer

calibration will not be stored.

Note: The replaced buffer is not removed from the calibration list and

can be selected for the next calibration.

WORKING WITH CUSTOM BUFFERS
If a custom buffer was configured in SETUP menu, it can be selected for

calibration by pressing the ARROW keys. The “CUSTOM” tag will be

displayed blinking on the LCD.
Press 2nd then Setup if you want to adjust

the buffer value to it’s value at the current

temperature. The buffer value, displayed on

the secondary LCD, will start blinking.
Use the ARROW keys to change the buffer value.
After about 5 seconds, the last buffer change

will be updated. Press 2nd then Setup if you want to change it again.
Notes: • Custom buffer value can be adjusted in a ±1.00 pH window,

around the set value.

• If you want to return to the set custom buffer value, simply

press the ARROW keys ( then ) to select again the

custom buffer.