beautypg.com

Hanna Instruments HI 98160 User Manual

Page 13

background image

13

• The calibrated value is then displayed on the primary LCD and the

secondary LCD will display the fifth expected buffer value.

• After the fourth calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into

the fifth buffer solution and stir gently. The temperature probe

should be close to the pH electrode.

• If necessary, press the ARROW keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected buffer, “CFM”

tag blinks.

• Press CFM to confirm calibration.

• The instrument stores the calibration values and

returns to normal measurement mode.

Note: For HI 98140 calibration proceed as in the FIVE-POINT CALIBRATION

above. After the third point is confirmed the instrument stores the

calibration value and returns to measurement mode.

FOUR, THREE or TWO-POINT CALIBRATION
• Proceed as described in FIVE-POINT CALIBRATION section.

• Press CAL after the appropriate accepted calibration

point. The instruments will return to measurement

mode and will memorize the calibration data.

ONE-POINT CALIBRATION
Two SETUP selectable options are available for one-point calibration:

“Pnt” and “OFFS”. See page 22 for Setup information using “1Pnt”

parameter.

If the “Pnt” option is selected, the adjacent slopes will be recalculated

using the new calibration value.

If the “OFFS” option is selected, an electrode offset correction is performed

keeping the existing slopes unchanged, but offsetting all by same value.

• Proceed as described in FIVE-POINT CALIBRATION section.

• Press CAL after the first calibration point was confirmed. The

instruments will memorize the one-point calibration

data and will return to measurement mode.

Notes: • Press Range to toggle between pH buffer

and temperature reading during calibration.

• If the value measured by the instrument is not close to the

selected buffer, “WRONG” “ ” and “WRONG” “ ” tags

will blink alternately. In this case check if the correct buffer

has been selected or used, or regenerate the electrode by

following the Cleaning Procedure (see page 39). If necessary,

change the buffer or the electrode.

• If the buffer temperature or the manual temperature exceeds

the temperature limits of the buffer, “WRONG” “ ” tags

and temperature reading will blink.