Note: The custom buffers can be set only with 0.01 pH resolution, at

25 ºC. Press Range (HI 98150 & HI 98160) or 2nd and Mode
(HI 98140) repeatedly while in custom buffer setting mode until
the closest buffer value to the desired custom buffer to be set is
displayed on the LCD. If selecting “no” option, the selected custom
buffer is removed. After removing one custom buffer, the custom
buffer list is reordered.

This feature allows the user to log pH or Rel mV measurements, together
with corresponding mV and temperature automatically. All logged data
can be transferred to a PC through the USB port.
The maximum logging space is 500 record locations.
LOGGING THE CURRENT DATA
To store the current reading into memory, press
Log while in measurement mode.
The instrument will display the current date
(mm.dd) on the primary LCD, the record number
on the secondary LCD and then the available
number of logs remaining.

If there are less than 6 memory locations
remaining, the “Lo” message will be displayed
for a few seconds to alert the user, and then
the available number of remaining logs is
displayed on the LCD.

If the LOG space is full, “FULL LOG” message
will be displayed on the LCD for a few seconds
and then ”FrEE 0” message.
The instrument returns to normal measurement
mode.