Hanna Instruments pH 210 User Manual

Page 13

background image

24

25

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

pH BUFFER SOLUTIONS

HI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcs
HI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcs
HI 70010P pH 10.01 Buffer Sachets, 20 mL, 25 pcs
HI 7004L pH 4.01 Buffer Solution, 500 mL bottle
HI 7006L pH 6.86 Buffer Solution, 500 mL bottle
HI 7007L pH 7.01 Buffer Solution, 500 mL bottle
HI 7009L pH 9.18 Buffer Solution, 500 mL bottle
HI 7010L pH 10.01 Buffer Solution, 500 mL bottle
HI 8004L pH 4.01 Buffer Sol. in FDA approved bottle, 500 mL
HI 8006L pH 6.86 Buffer Sol. in FDA approved bottle, 500 mL
HI 8007L pH 7.01 Buffer Sol. in FDA approved bottle, 500 mL
HI 8009L pH 9.18 Buffer Sol. in FDA approved bottle, 500 mL
HI 8010L pH 10.01 Buffer Sol. in FDA approved bottle, 500 mL

ELECTRODE STORAGE SOLUTIONS

HI 70300L Storage Solution, 460 mL bottle
HI 80300L Storage Solution in FDA approved bottle, 460 mL

ELECTRODE CLEANING SOLUTIONS

HI 70000P Electrode Rinse Sachets, 20 mL, 25 pcs
HI 7061L General Cleaning Solution, 460 mL bottle
HI 7073L Protein Cleaning Solution, 460 mL bottle
HI 7074L Inorganic Cleaning Solution, 460 mL bottle
HI 7077L Oil & Fat Cleaning Solution, 460 mL bottle
HI 8061L General Cleaning Sol. in FDA approved bottle, 460 mL
HI 8073L Protein Cleaning Solution in FDA approved bottle, 460 mL
HI 8077L Oil & Fat Cleaning Sol. in FDA approved bottle, 460 mL

ELECTRODE REFILL ELECTROLYTE SOLUTIONS

HI 7071

3.5M KCl+AgCl Electrolyte, 4x50 mL, for single junction
electrodes

HI 7072

1M KNO

3

Electrolyte, 4x50 mL

HI 7082

3.5M KCl Electrolyte, 4x50 mL, for double junction
electrodes

HI 8071 3.5M KCl+AgCl Electrolyte in FDA approved bottle,

4x50 mL, for single junction electrodes

HI 8072 1M KNO

3

Electrolyte in FDA approved bottle, 4x50 mL

HI 8082 3.5M KCl Electrolyte in FDA approved bottle, 4x50 mL

Typical Electrode Life

Ambient Temperature

1- 3 years

90 °C

Less than 4 months

120 °C

Less than 1 month

Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline

solutions. The pH at which the interference starts to be significant

depends upon the composition of the glass. This interfe-rence is called

alkaline error and causes the pH to be underestimated. Hanna’s glass

formulations have the indicated characteristics.

1.0 Mol L

-1

Na

+

0.1 Mol L

-1

Na

+

Sodium Ion Correction for Glass at 20-25 °C

Concentration

pH

Error

13.00

13.50

14.00

12.50

13.00

13.50

14.00

0.10

0.14

0.20

0.10

0.18

0.29

0.40

The resistance of glass electrodes partially depends on the temperature.

The lower the temperature, the higher the resistance. It takes more time

for the reading to stabilize if the resistance is higher. In addition, the

response time will suffer to a greater degree at temperatures below 25 °C.

Since the resistance of the pH electrode is in the range of 50-200 Mohms,

the current across the membrane is in the pico Ampere range. Large

currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and

static discharges are detrimental to a stable pH reading.
The pH electrode’s life also depends on the temperature. If constantly

used at high temperatures, the electrode life is drastically reduced.

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

FOR

FOR

FOR

FOR

FOR pppppH SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS