ENMET EX-5000 User Manual

Page 7

background image

Catalytic S/T

ENMET Corporation

5

Figure 3: Catalytic Combustible Gas Sensor/Transmitter

1. Apply the calibration gas associated with the S/T to the sensor. See to Figure 3.

2. Allow the gas to flow 3 minutes.

3. Use the magnet to adjust the span potentiometer up or down, until the display reads the calibration gas value.

4. Remove the calibration gas.

5. In clean air, adjust the zero point if necessary, using the magnet to adjust the zero potentiometer.

6. If the zero point adjustment is large, greater than 10% LEL, apply the calibration gas again, and make a second

span adjustment.

7. Calibration is complete.

2.3 Standard Range and Alarm LED

The standard range is 0-100%LEL of the combustible gas being monitored, with 100%LEL producing 20mA in the
loop. The red alarm LED is set to turn on at the high alarm level of the gas. The high alarm level and particular gas
are indicated above the LED.

Red ‘Alarm’ LED

Zero +

Zero

––

Span

––

Span +

Calibration Cover 03620-012
use with sensor 03070-001

Calibration Cover 03620-015

use with sensor 03070-003

Fitting to
calibration cover

Regulator

Gas Cylinder