0 operation – ENMET EX-2000 User Manual

Page 8

background image

ENMET Corporation

EX2000

4

3.0 Operation

3.0 Operation

3.0 Operation

3.0 Operation

Refer to Figure 1 for location of the

EX-2000

switches, alarms and display. While in operation the

EX-2000

will regularly

sound a chirp or “beep” to indicate the instrument is turned on and in normal operation with no faults detected. This confidence
chirp is set at the factory and can not be altered or turned off.

EX-2000

is programmed at the factory for a specific target gas which is displayed on a label on the left side of the instrument.

C

AUTION

:

The

EX-2000

must be returned to

ENMET

or an authorized distributor to alter the target gas

The

ENTER

switch is used for both start-up and shutdown of the

EX-2000

. See section 3.1 for start-up & operation and section

3.2 for shutdown of the

EX-2000

instrument.

3.1 Turning On the EX2000

To start-up the

EX-2000,

press and release the

E

NTER

switch.

When the

EX-2000

is turned on the following self test and count down for sensor to stabilization in succession on the display,

accompanied by a continuous visual and audible signal.

Example of start-up sequence:

Display

Sequence

Audio and Visual alarms are active during this sequence:

1.

Whole display is flashed.
(4 large digits will flash, 3 large digits will display in normal
operation)

2.

Software version is displayed.

3.

Alarm 1 threshold is displayed.

4.

Alarm 2 threshold is displayed.

5.

Self-test is indicated.

6.

Countdown of sensor stabilization time is displayed.
(10 seconds maximum)

7.

Operational measurement displayed.
And Audio & Visual Alarms go to the normal state.

A regular audible “beep” signals that the instrument is operating correctly. The interval between the “beeps” is factory-
programmable and the “beeps” can not be disabled.

r 1.6

2

TST

1

TST

0

TST