4.2 Calibrating the EX-2000

The only way to check if the

EX-2000

detection function is working properly is the calibration performed with a standard gas.

Calibration must be performed:

Every 6 months minimum.

After an exposure to high gas concentrations.

After more than one month storage with no use.

After a sensor change.

Following an instrument fault.

The recommended standard gas is stated on the target gas label, located on the side of the

EX-2000

. Other standard gases may

be used. See Table 1, Coefficients of gases detected.

To enter the Calibration Menu:

1.

Enter the calibration menu by inserting the programming plug into the programming port and pressing the

ALARM

&

BACKLIGHT

switches simultaneously. The pictogram of the maintenance wrench will appear and PrOG will

appear on the display, to indicate the

EX-2000

is in maintenance mode and ready for programming or calibration. Press

the

BACKLIGHT

switch to toggle between programming and calibration.

2.

While CAL is displayed press the

ENTER

switch to enter the calibration menu.

While Set0 is displayed press the

ENTER

switch to adjust the zero value.

3.

Modify the zero value if necessary:

Press the

BACKLIGHT

switch to decrease the zero value. ▼

Press the

ALARM

switch to increase the zero value. ▲

4.

Press the

ENTER

switch to enter the new zero value.

SEnS will appear on the display, for the adjustment of sensitivity to standard gas.

5.

Press the

ENTER

switch to enter the sensitivity adjustment procedure.

6.

Attach the calibration cover to the calibration adapter and position the calibration cover over the sensor.

7.

Attach the standard gas to the calibration adapter and open valve on cylinder of gas to allow the gas to flow

The measurement of applied standard gas will appear on the display, for the adjustment of sensitivity to standard gas.
Allow signal to stabilize, approximately 2 minutes.

8.

Modify the sensitivity value if necessary:

Press the

BACKLIGHT

switch to decrease the sensitivity value. ▼

Press the

ALARM

switch to increase the sensitivity value. ▲

9.

Press the

ENTER

switch to request confirmation of the new sensitivity value.

COnF will appear on the display for the confirmation of the new sensitivity value.

10.

Press the

ENTER

switch to confirm of the new sensitivity value.

COnFNO will appear on the display, to request rejection or confirmation of sensitivity value.

Press the

BACKLIGHT

switch to toggle between Confirmation Yes or No.(COnFYES or COnFNO)

11.

Press the

ENTER

switch to confirm or reject the new sensitivity value.

The instrument will return to the Operational Mode.

Remove calibration adapter and cover.

Remove the programming plug.