4.2 Calibrating the EX-2000
The only way to check if the
EX-2000
detection function is working properly is the calibration performed with a standard gas.
Calibration must be performed:
Every 6 months minimum.
After an exposure to high gas concentrations.
After more than one month storage with no use.
After a sensor change.
Following an instrument fault.
The recommended standard gas is stated on the target gas label, located on the side of the
EX-2000
. Other standard gases may
be used. See Table 1, Coefficients of gases detected.
To enter the Calibration Menu:
1.
Enter the calibration menu by inserting the programming plug into the programming port and pressing the
ALARM
&
BACKLIGHT
switches simultaneously. The pictogram of the maintenance wrench will appear and PrOG will
appear on the display, to indicate the
EX-2000
is in maintenance mode and ready for programming or calibration. Press
the
BACKLIGHT
switch to toggle between programming and calibration.
2.
While CAL is displayed press the
ENTER
switch to enter the calibration menu.
While Set0 is displayed press the
ENTER
switch to adjust the zero value.
3.
Modify the zero value if necessary:
Press the
BACKLIGHT
switch to decrease the zero value. ▼
Press the
ALARM
switch to increase the zero value. ▲
4.
Press the
ENTER
switch to enter the new zero value.
SEnS will appear on the display, for the adjustment of sensitivity to standard gas.
5.
Press the
ENTER
switch to enter the sensitivity adjustment procedure.
6.
Attach the calibration cover to the calibration adapter and position the calibration cover over the sensor.
7.
Attach the standard gas to the calibration adapter and open valve on cylinder of gas to allow the gas to flow
The measurement of applied standard gas will appear on the display, for the adjustment of sensitivity to standard gas.
Allow signal to stabilize, approximately 2 minutes.
8.
Modify the sensitivity value if necessary:
Press the
BACKLIGHT
switch to decrease the sensitivity value. ▼
Press the
ALARM
switch to increase the sensitivity value. ▲
9.
Press the
ENTER
switch to request confirmation of the new sensitivity value.
COnF will appear on the display for the confirmation of the new sensitivity value.
10.
Press the
ENTER
switch to confirm of the new sensitivity value.
COnFNO will appear on the display, to request rejection or confirmation of sensitivity value.
Press the
BACKLIGHT
switch to toggle between Confirmation Yes or No.(COnFYES or COnFNO)
11.
Press the
ENTER
switch to confirm or reject the new sensitivity value.
The instrument will return to the Operational Mode.
Remove calibration adapter and cover.
Remove the programming plug.