ENMET EN Series User Manual
Page 17

Sensor/Transmitter, Toxic Gas & Oxygen
ENMET Corporation
15
Phosgene (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
ammonia
50
5, drops to 0 in short time
abundant change in humidity
yes
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen sulfide
1*
sulfur dioxide
2
hydrochloric acid
5*
carbon monoxide
300
carbon dioxide
5000
nitrogen
100%
chlorine
1
hydrocarobons
% range
hydrogen
1000
*With onboard filter; continuous high level exposure may reduce the filter efficiency
Silane (3E 50 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine
0,5
-4
hydrogen cyanide
10
1
ammonia
100
<1
diborane
100
0.4
arsine
1
0.7
germane
1
1
disilane
5
yes
phosphine
300ppb
0.2
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen sulfide
1*
sulfur dioxide
2
hydrochloric acid
5
‡
carbon monoxide
300
carbon dioxide
5000
nitrogen
100%
chlorine
1,5
hydrocarobons
% range
hydrogen
1000
hydrogen fluoride
3.8
hydrogen selenide
50ppb
*With onboard filter; continuous high level exposure may reduce the filter efficiency
‡Short term gas exposure (min.)
Notes: