Appendix a: interference table – ENMET EN Series User Manual
Page 10
Sensor/Transmitter, Toxic Gas & Oxygen
ENMET Corporation
8
Appendix A: Interference Table
For Selected Sensors
Ammonia (3E-100 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
ethanol
1000
1
isopropanol
450
1
hydrogen sulfide
14
10
methanol
1200
3
hydrogen
1000
80
carbon monoxide
300
100
sulfur dioxide
25
-21
nitrogen dioxide
50
25
hydrogen cyanide
10
-18
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
chlorine
5
unsaturated hc (ethylene)
1.0 %
saturated hydrocarbons
abundant
Carbon Monoxide (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
hydrogen
1000
450
nitric oxide
100
25
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
ammonia
100
chlorine
5
hydrogen cyanide
10
ethylene
2.0 % *
carbon dioxide
5,000
methane
10,000
sulfur dioxide
10 *
hydrogen sulfide
10 *
nitrogen dioxide
10 *
isopropanol
1,025 *
gasoline vapor
saturated *
* With onboard filter; continuous high level exposure may reduce the filter efficiency