Sensor/Transmitter, Toxic Gas & Oxygen
ENMET Corporation
10
Hydrogen Cyanide(2E 30 sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
hydrogen sulfide
15
40
†
chlorine
5
-1*
nitrogen dioxide
15
-40
†
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen chloride
10*
methane
2000
carbon monoxide
300
carbon dioxide
5,000
gasoline
300
sulfur dioxide
50*
nitric oxide
100
chlorinated HC's
<200
†Long term exposure may destroy sensor
*with onboard filter; continuous high level exposure may reduce the filter efficiency