11 menu mode – ENMET EX-6100 Series User Manual
Page 19
![background image](/manuals/736163/19/background.png)
EX-6100
ENMET Corporation
16
11 Menu Mode
The
EX-6100
uses a password system to restrict the end user from carrying out certain changes that may compromise the use of the
equipment. Menu Mode should only be entered by qualified personal.
To enter the menu mode Sequence Tap:
Menu, Up, Down, Enter, Up, Enter, Up, Menu
If no inputs (taps) the
EX-6100
returns to operational mode.
Several calibration modes exist in the
EX-6100
and these are accessible via the instruments simple menu system. To select a calibration
mode follow this procedure:
•
Press the MENU button, E: 1 appears on the display.
•
Press UP or DOWN until the required menu option is displayed.
•
Press ENTER to select the calibration mode.
•
To exit the menu press MENU.
While the instrument is in menu mode – any data displayed on the screen will alternate between the cal number and the reading.
The following features are available via the ‘EX-6100’ menu system:
Table 3: EX-6100 Menu Options
Menu Option
Section
E : 1 – Sensor Zero
11.2.1
E : 2 – Sensor Span
11.2.2
E : 3 – Sensor FSD
11.2.3
E : 4 – Output Zero (4mA)
11.2.4
E : 5 – Output Span (20mA)
11.2.5
E : 6 – Decimal Places
11.2.6
E : 7 – Firmware Version
11.2.7
E : 8 – Restore
11.2.8
E : 9 – Engineer Diagnostics
11.2.9
E : 12 – Cross Reference
11.2.10
E : 13 – Relay 1
11.2.11
E : 14 – Relay 2
11.2.12
E : 15 – Alarm Level 1
11.2.13
E : 16 – Alarm Level 2
11.2.14
E : 19 – Positive Zero Suppression
11.2.15
E : 20 – Negative Zero Suppression
11.2.16
E : 27 – Hysteresis
11.2.17
E : 28 – Firmware Update
11.2.18
E : 29 – Relay 3
11.2.19
E : 30 – Password
11.2.20
E : 31 – Warm-Up Timer
11.2.21
Contact ENMET Corporation before entering this menu option.
Do Not enter this menu option.
11.2.1 E : 1 – Sensor Zero
Refer to section 5.1.1 for sensor calibration details.
11.2.2 E : 2 – Sensor Span
Refer to section 5.1.2 for sensor calibration details.
11.2.3 E : 3 – Sensor FSD
This menu option allows the full-scale gas level to be selected.
From the menu system select menu option: E: 3 and press ENTER.
•
Using the INCREASE and DECREASE buttons; adjust the FSD to the required level.
•
Press ENTER, ‘- - - - ‘will be displayed to indicate that the new setting has been accepted.
Note: Pressing MENU instead of ENTER will exit the menu without changing the FSD.
•
Press MENU to close the menu system.
11.2.4 E : 4 – Output Zero (4mA)
Refer to section Appendix A for output calibration details.
11.2.5 E : 5 – Output Span (20mA)
Refer to section Appendix A for output calibration details.