SPECTRUM SP Series

ENMET Corporation

15


Table 2: SPECTRUM SP Calibration Voltage and Countdown Times

N

OTE

:

The following are values used for production calibration. Not all gases are available for field calibration; contact

ENMET

customer service personnel.

Gas

Calibration Gas

Calibration Voltage

Countdown Time

Ammonia

25ppm

0.67

180 sec

Arsine

0.5ppm PH

3

= 0.38ppm

0.10

120 sec

Carbon Monoxide

100ppm

1.70

120 sec

Chlorine

5ppm

1.40

180 sec

Fluorine

3ppm Cl

2

= 7.5ppm

1.80

240 sec

Hydrazine

0.3ppm

0.91

240 sec

Hydrogen

800ppm

2.20

180 sec

Hydrogen Chloride

10ppm

2.10

180 sec

Hydrogen Cyanide

10ppm

2.10

120 sec

Hydrogen Fluoride

5ppm

1.40

240 sec

Hydrogen Sulfide

20ppm

0.46

120 sec

Nitric Oxide

25ppm

0.67

90 sec

Nitrogen Dioxide

5ppm

1.30

120 sec

Oxygen

20.9%(air)

2.80

120 sec

Ozone

0.3ppm

0.75

240 sec

Phosgene

0.66ppm

0.18

240 sec

Phosphine

0.5ppm

0.13

120 sec

Silane

5ppm

0.54

120 sec

Sulfur Dioxide

10ppm

2.10

90 sec

N

OTE

:

All specifications stated in this manual may change without notice.