Appendix A: Calibration Data Tables
Table 1: Gas Ranges, Alarm Points and Sensor Life
Gas
Range
Alarm
lower limit
Alarm
upper limit
Factory Alarm
Set Point
Expected Sensor
Lifetime
Ammonia
0-100ppm
15ppm
50ppm
25ppm
1 year
Arsine
0-0.99ppm
0.03ppm
0.10ppm
0.05ppm
1-1.5 years
Carbon Monoxide
0-1000ppm
5ppm
200ppm
35ppm
2-3 years
Chlorine
0-10ppm
0.5ppm
5.0ppm
1.0ppm
1-2 years
Fluorine
0-10ppm
1ppm
5ppm
1ppm
2-3 years
Hydrogen
0-1000ppm
200ppm
750ppm
200ppm
1-2 years
Hydrogen Chloride ① 0-20ppm
3.0ppm
10.0ppm
5.0ppm
1-2 years
Hydrogen Cyanide
0-20ppm
3.0ppm
10ppm
4.7ppm
2-3 years
Hydrogen Fluoride
0-10ppm
2.0ppm
6.0ppm
3.0ppm
1 year
Hydrogen Sulfide
0-200ppm
5.0ppm
25ppm
10ppm
2-3 years
Nitric Oxide ①
0-100ppm
15ppm
50ppm
25ppm
1-2 years
Nitrogen Dioxide
0-10ppm
1.0ppm
5.0ppm
3.0ppm
1-2 years
Ozone
0-0.99ppm
0.05ppm
0.2ppm
0.1ppm
1-1.5 years
Phosgene
0-0.99ppm
0.05ppm
0.20ppm
0.1ppm
1 years
Phosphine
0-0.99ppm
0.15ppm
0.30ppm
0.50ppm
1-1.5 years
Silane
0-30ppm
3ppm
5ppm
5ppm
1-1.5 years
Sulfur Dioxide
0-20ppm
2.0ppm
10ppm
5ppm
1-2 years
N
OTE
:
All specifications stated in this manual may change without notice.
① Biased Sensors
:
Instruments using biased sensor, stabilization time is extended to 4 minutes. Complete stabilization may
take as long as 1 hour. When instrument is turned on it may display
XXXX
+
. If instrument has been off
for an extended amount of time it may be necessary to “cycle” (turn on and off) the instrument several times.
See Appendix C for Hydrazine