ENMET SPECTRUM SP User Manual
Page 11

SPECTRUM SP Series
ENMET Corporation
9
3.1.4 Changing the Alarm Level
A valid key entry sets the instrument at the "run ZERO" location of the maintenance menu. Press the
OPTION
pushbutton
twice to access the "set ALM" display, then press the
SELECT
pushbutton; ALM=0000" is displayed. This is called the alarm
update window, and the value displayed is the present alarm setpoint. The underscore cursor is under the far left digit. Press
the
OPTION
pushbutton to change the underscored digit; select the desired digit, and press the
SELECT
pushbutton to lock in
the desired digit and move the underscore cursor one position to the right. When the desired new alarm point is set
“new=XXXX” is displayed. Press either the
OPTION
or
SELECT
pushbutton to exit the alarm update window. If the new
alarm setting is valid, "exit" is displayed. Press the
OPTION
pushbutton to return to the operational display, or the
SELECT
pushbutton to return to “run ZERO”.
For the safety of the user, there are upper and lower limits past which the alarm setting is invalid, and the instrument does not
accept them. If an invalid alarm setting is attempted, after the numerical value is inserted in the "ALM=0000" window,
pressing the
OPTION
or
SELECT
pushbutton results in a momentary display of "INVALID" after which the display returns to
the alarm update window. Exiting the alarm update window at this point results in an alarm point setting unchanged from the
value present when the procedure was begun.
Factory default alarm setpoints and alarm limits are shown in Table 1 Appendix A.
3.1.5 Setting a New Key
A valid key entry sets the instrument at the "run ZERO" location of the maintenance menu. Press the
OPTION
pushbutton four
times to access the "set KEY" display. Press the
SELECT
pushbutton once; "KEY=0000 is displayed. A new key can be set
by changing the underscored number with the OPTION pushbutton and moving the underscore cursor with the
SELECT
pushbutton. After the new key is entered “new-XXXX” is displayed, press the
OPTION
or
SELECT
pushbutton to display to
“exit”, then press the
OPTION
pushbutton to return to “run ZERO”.
N
OTE
:
Four digit key numbers should be selected carefully and recorded. Without the correct key, the maintenance menu
cannot be accessed. If a four digit key number is lost, call
ENMET
customer service personnel.