beautypg.com

Figure 3 quadrant maintenance menu flow chart – ENMET QUADRANT User Manual

Page 9

background image

QUADRANT

ENMET Corporation

3

LEL
E

Scrolling display of the gas measurements

Figure 3

QUADRANT maintenance menu flow chart

Combustible (CH4)

O2

CO

H2S

From figure 2 page 2

Cod 0000
(password)

E

Auto-Zero

E yes

E done

(Auto-Set)

no

+

+

Program

+

Calibrate(calib) +

Cell Change + End +

E

E

E

E

chan

chan

chan

XXX

XXX

XXX
E

E

E

CG
XXX

ON/OFF

E

E

0 settin

GAS CG

E

(CH4)

XXX

YY:XX

E

E

E

Ran
XXX

0: XXX

E

E

S settin
E

AL XX

S: XXX

YY:XX

E

E

E

Yes/No Yes/No

Yes/No

E = Yes

E = Yes

E = Yes

= No

= No

= No

+ to change channel

+ to put channel in on or
off mode

+ to change gas
(combustible)

Change alarm point
+ to increment digit
– to move to next digit
O2 has high & low
alarms

+ to change
channel

To change cal gas
value (except O2)
+ to increment
to move

Wait for the
reading to stabilize
(except O2)

Apply calibration
gas wait for the
reading to stabilize

combustible only
+ to change
combustible gas
display

To input preset password
+ to increment digit
to move to next digit

+ to change channel

To change cal gas
value (except O2)
+ to increment
to move

Wait for general LED
to turn off
(except O2)

Apply calibration
gas wait for the
reading to stabilize

Apply calibration gas
and wait for general
LED to turn off

Maximum range of
channel

YY = position of
digital pot
XX = sensor reading