3.7 Alarm Information
Alarm type
Information displayed
Low alarm:
Slow Pulsed tone alarm sound
Alarm LED Flashing
Vibration
High alarm:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Vibration
STEL alert:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Vibration
TWA alert:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Vibration
Overrange alert:
Quick Pulsed tone alarm sound
Alarm LED Flashing
Battery low voltage alert:
log appears.
At this time, the detector can still work for 30
days at least. When the battery is used up, the
detector will turn off automatically.
Sensor end-of-life indication:
Within 0-9 days before the sensor is used up,
screen will display as the right picture shows
when the detector is self testing after turning on.
The figure means the number of the days
remaining.
Table 1: RECON-IS Range and Alarm Levels
Target gas
Detecting Range
Low alarm level
High alarm level
TWA level
STEL level
H
2
S 0-100ppm
10ppm 15ppm
10ppm
15ppm
CO 0-1000ppm
35ppm 100ppm
25ppm
200ppm
O
2
0-30%vol
19.5%vol
23.5%vol
-- --