3.7 Alarm Information

Alarm type

Information displayed

Low alarm:

 Slow Pulsed tone alarm sound

 Alarm LED Flashing

 Vibration

High alarm:

 Quick Pulsed tone alarm sound

 Alarm LED Flashing

 Vibration

STEL alert:

 Quick Pulsed tone alarm sound

 Alarm LED Flashing

 Vibration

TWA alert:

 Quick Pulsed tone alarm sound

 Alarm LED Flashing

 Vibration

Overrange alert:

 Quick Pulsed tone alarm sound

 Alarm LED Flashing

Battery low voltage alert:

log appears.

 At this time, the detector can still work for 30

days at least. When the battery is used up, the
detector will turn off automatically.

Sensor end-of-life indication:

 Within 0-9 days before the sensor is used up,

screen will display as the right picture shows
when the detector is self testing after turning on.
The figure means the number of the days
remaining.

Table 1: RECON-IS Range and Alarm Levels

Target gas

Detecting Range

Low alarm level

High alarm level

TWA level

STEL level

H

2

S 0-100ppm

10ppm 15ppm

10ppm

15ppm

CO 0-1000ppm

35ppm 100ppm

25ppm

200ppm

O

2

0-30%vol

19.5%vol

23.5%vol

-- --