Duration: Duration” determines the total process time since control

start.

The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be

set to ”Off”. (”Off” means that no endpoint for pump down is preset.)

The screen-shot shows the factory-set values.

detect
Sensitivity normal

Hysteresis

Auto

Delay

Off

Duration

Off

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

00:00:00

When selecting ”Graphic” the display shows a pressure vs. time curve.

The timeline in the diagram adapts automatically to the process time.
+

Press the selection knob twice to return to the standard display.

430 mbar

00:07:45

detect

+

When pressing ”STOP” key, or after process stop due to elapsed ”Du-

ration” (preset process time), the controller is set again to the ”detect”

function.

Temporary switching from ”detect” to ”Vac control” (only if control

is running):
+

Press ”MODE” key. The controller switches to ”Vac control” function,

the current vacuum is used as set value.

+

The preset function of the controller does not change due to this tem-

porary switching. When pressing ”STOP” key, the controller is set

again to the ”detect” function.

Attention:

The ”detect” function is designed for operation with a stand-alone vac-

uum pump. If the CVC 3000 detect is to be operated in a vacuum

network you must ensure that during the determination of the boiling

vacuum no disturbance due to other users will occur.