VACUUBRAND CVC 3000 detect User Manual
Page 39
+
Duration: ”Duration” determines the total process time since control
start.
The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be
set to ”Off”. (”Off” means that no endpoint for pump down is preset.)
The screen-shot shows the factory-set values.
detect
Sensitivity normal
Hysteresis
Auto
Delay
Off
Duration
Off
- - - - - - Graphic - - - - - -
- - - - - - - Back - - - - - - -
00:00:00
When selecting ”Graphic” the display shows a pressure vs. time curve.
The timeline in the diagram adapts automatically to the process time.
+
Press the selection knob twice to return to the standard display.
430 mbar
00:07:45
detect
+
When pressing ”STOP” key, or after process stop due to elapsed ”Du-
ration” (preset process time), the controller is set again to the ”detect”
function.
Temporary switching from ”detect” to ”Vac control” (only if control
is running):
+
Press ”MODE” key. The controller switches to ”Vac control” function,
the current vacuum is used as set value.
+
The preset function of the controller does not change due to this tem-
porary switching. When pressing ”STOP” key, the controller is set
again to the ”detect” function.
➨
Attention:
The ”detect” function is designed for operation with a stand-alone vac-
uum pump. If the CVC 3000 detect is to be operated in a vacuum
network you must ensure that during the determination of the boiling
vacuum no disturbance due to other users will occur.