VACUUBRAND MV 2 VARIO User Manual
Page 36
![background image](/manuals/715242/36/background.png)
page 36 of 41
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999035 / 15/12/2009
Montage of the connecting hose
If the pump does not achieve the ultimate pressure:
☞ In case the diaphragms and valves have been replaced, a run-in period of several
hours is required before the pump achieves its ultimate vacuum.
If all pump heads achieve a vacuum below 110 mbar but pump does not achieve the
ultimate total pressure:
Check hose connectors between pump heads and manifolds for leaks. If necessary
recheck pump chamber.
➨ Use open-ended wrench to assemble fittings with connec-
tion hoses to all pump heads.