Notice – VACUUBRAND MV 10 User Manual
Page 37
page 37 of 51
Ensure that maintenance is done only by suitably
trained and supervised technicians.
The valves and diaphragms as well as the motor capac-
itors are wear parts. If the rated ultimate vacuum is no
longer achieved or in case of increased noise level, the
pump interior, the diaphragms and the valves must be
cleaned and the diaphragms and valves must be checked
for cracks or other damage.
All bearings are encapsulated and are filled with long-life
lubricant. Under normal operating conditions, the drive
system is maintenance free.
In demanding circumstances, it may be efficient to
check and clean the pump heads on a regular basis. In
normal use, the lifetime of the diaphragms and valves
is typically 15,000 operating hours.
- Prevent internal condensation, transfer of liquids or
dust. The diaphragms and valves will be damaged if
liquid is pumped in significant amount.
- Carry out maintenance frequently if the pump is ex-
posed to corrosive media or in case of deposits.
- Regular maintenance will improve the lifetime of the
pump and also protect both users and the environment.
NOTICE
Set of seals for ME 16, MD 8 (diaphragms and valves) ..............................696819
Set of seals for MD 12, MV 10 (diaphragms, valves and O-rings)
696827
16 x O-ring 28 x 1.5 (MD 8 / MD 12 / MV 10) ..............................................637090
Diaphragm key (w/f 66) ...............................................................................636554
Tools required (metric):
- Phillips screwdriver size 2
- Open end wrench width 15 / 17 / 20
- 5 mm wide Allen key
- Diaphragm key width 66 mm