Maintenance and repair, Notice – VACUUBRAND MV 10 User Manual
Page 19
page 19 of 51
In case of overload, the motor is shut down by a
self-hold
thermal cutout in the winding.
Three-phase motors are equipped with a circuit- breaker
which isolates all poles in case of an overload.
Note: Only manual reset is possible. Switch off the pump
and disconnect from the power source. Identify and elimi-
nate the cause of failure. Allow the pump to cool down suf-
ficiently before restart. Wait approx. five minutes before
restarting the pump.
Maintenance and repair
In order to comply with laws (occupational, health and
safety regulations, safety at work law and regulations for
environmental protection) vacuum pumps, components
and measuring instruments can only be returned when
certain procedures (see section „Notes on return to the
factory“, pg. 45) are followed.
Take advantage of our service seminars, which put special
focus on the maintenance and repair of vacuum pumps.
For details and for the online ”Instructions for repair” man-
In normal use, the lifetime of the diaphragms and valves
is typically 15,000 operating hours. Bearings have a typi-
cal durability of 40000 h. Motor capacitors have a typi-
cal durability in the range of 10000 to 40000 h depending
strongly on operation conditions including ambient tem-
perature, humidity or load.
➨
Ensure that the pump cannot be operated acciden-
tally. Never operate the pump if covers or other
parts of the pump are disassembled.
➨
Switch off the pump. Disconnect the electrical
power cord and wait five seconds before starting
maintenance to allow the capacitors to discharge.
NOTICE