VACUUBRAND MV 10 User Manual
Page 35
page 35 of 51
➨
A service manual with exploded view drawings, spare parts list and
directions for repair is available on request.
+
The service manual is intended for trained service people only.
Fault
Possible cause
Remedy
❑ Pump too noisy.
➨ Atmospheric or high
pressure at the pump
inlet?
✔ Connect hose or silencer
to pump outlet. Be careful
not to cause outlet over-
pressure, especially with
condensable vapors.
➨ Diaphragm crack or
diaphragm clamping
disc loose?
✔ Perform maintenance.
➨ Other than above men-
tioned causes?
✔ Contact local distributor.
❑ Pump seized.
✔ Contact local distributor.