Read commands ”cvc 3000, Read commands ” cvc 3000 – VACUUBRAND PC 3012 NT VARIO Duo User Manual
Page 93
page 87 of 145
Read commands ”CVC 3000”
Command
Operation
Response
Description
IN_PV_1
current pressure XXXX.X mbar/Torr/hPa unit according to preselections
IN_PV_2
current speed
XXX%
1-100% or ”HI”
IN_PV_3
time
XX:XX h:m
process runtime (hours:minutes)
IN_PV_X
pressure
XXXX.X XXXX.X ...
pressure of all connected sensors,
unit (mbar/Torr/hPa) according to
preselections
IN_PV_T
operation time of
the controller
XXXXdXXh
operation time in days and hours
IN_CFG
device set
preselections
0XXXXXXXXXXXXXXX
1XXXXXXXXXXXXXXX
2XXXXXXXXXXXXXXX
3XXXXXXXXXXXXXXX
4XXXXXXXXXXXXXXX
5XXXXXXXXXXXXXXX
XyXXXXXXXXXXXXXX
XX0XXXXXXXXXXXXX
XX1XXXXXXXXXXXXX
XX2XXXXXXXXXXXXX
XXX0XXXXXXXXXXXX
XXX1XXXXXXXXXXXX
XXXX0XXXXXXXXXXX
XXXX1XXXXXXXXXXX
XXXXX0XXXXXXXXXX
XXXXX1XXXXXXXXXX
XXXXXX0XXXXXXXXX
XXXXXX1XXXXXXXXX
XXXXXXX0XXXXXXXX
XXXXXXX1XXXXXXXX
XXXXXXXX0XXXXXXX
XXXXXXXX1XXXXXXX
XXXXXXXXX0XXXXXX
XXXXXXXXX1XXXXXX
XXXXXXXXXX0XXXXX
XXXXXXXXXX1XXXXX
XXXXXXXXXXX0XXXX
XXXXXXXXXXX1XXXX
XXXXXXXXXXXX0XXX
XXXXXXXXXXXX1XXX
XXXXXXXXXXXXXyXX
XXXXXXXXXXXXXXyX
XXXXXXXXXXXXXXX0
XXXXXXXXXXXXXXX1
VACUU•LAN
Pump down
Vac control
Auto mode
Program
measuring device
y: 0......D: language
+
(hexadecimal)
pressure unit mbar
pressure unit Torr
pressure unit hPa
autostart off
autostart on
acoustic signal off
acoustic signal on
VARIO pump not connected
VARIO pump connected
VMS not connected
VMS connected
in-line valve not connected
in-line valve connected
coolant valve not connected
coolant valve connected
venting valve not connected
venting valve connected
fault indicator not connected
fault indicator connected
level sensor not connected
level sensor connected
remote module not connected
remote module connected
y: 1......9: sensor number
y: 1......9: sensor quantity
remote operation off
remote operation on
+
Language:
0: German
1: English
2: French
3: Italian
4: Spanish
5: Turkish
6: Korean
7: Chinese
8: Portuguese
9: Russian
A: Polish
B: Dutch
C: Japanese
D: Finnish