beautypg.com

Shutdown & storage, Notice shutdown & storage, Notice – VACUUBRAND PC 3012 NT VARIO Duo User Manual

Page 55

background image

page 49 of 145

Drain catchpots.

Important: Comply with regulations when

disposing of solvents/condensates. Recy-

cle if possible; purify if contaminated.

Reattach drained catchpots.

NOTICE

Shutdown & storage

The pump can be switched off under vacuum.

Condensate?

O

➨ I

(5-10min)

atm

Short-term:

- Has the pump been exposed to conden-

sate?

Allow the pump to continue to run at atmo-

spheric pressure for a few minutes.

- Has the pump been exposed to media which

may damage the pump materials or form

deposits?

Check and clean pump heads if necessary.

- Has the pressure transducer been exposed

to media which may form deposits?

Clean pressure transducer if necessary.

Long-term:

- Take measures as described above regard-

ing short-term shutdown.

- Separate the pump from the application.

- Close inlet and outlet ports (e.g., with trans-

port caps).

- Close the gas ballast valve.

- Drain catchpots.

- Store the pump under dry conditions.

NOTICE

NOTICE