3
-Battery Life Warning Display
When the battery power becomes low, a battery life warning signal will appear in the
unit's LCD display.
If the battery power becomes even lower the unit will automatically go into sleep mode
in order to protect the data.
-15 Recording Intervals
Select from 15 recording intervals (from 1 second to 1 hour) to meet your needs.
Select from 2 Recording Modes
One-time Mode
ȇWhen the number of recorded readings reaches 8000, [FULL] will
appear in the unit's LCD display and recording will automatically stop.
Endless Mode
ȇWhen the number of recorded readings reaches 8000, the oldest data
readings will be overwritten and recording will continue.
-Backup Function
When battery power becomes low, a battery life warning signal will appear in the unit's
LCD display and if the battery power becomes even lower the unit will automatically go
into sleep mode in order to protect the data. In sleep mode all normal operations will
stop and it will become impossible to switch on the power of the main unit.
NOTE:
If the main unit remains in sleep mode for about 1 month without a change of battery,
or if the battery is left out of the unit for more than 2 minutes, all recorded data will
be lost.
-Current Readings Monitoring Display
With our exclusive software, you cannot only monitor the current measurements at a
set interval, but can view those measurements in a continually changing graph.
You can simultaneously display the current measurements and corresponding graphs
for the number of units you have connected.
-Adjustment Function
By entering the adjustment values beforehand with the provided software, it is possible
to view and record the adjusted measurement values.
There are two methods of adjustment: adjusting by one point or adjusting by two
points. Adjustment for differences will be based upon the following simple equation
Y=aX+b.
X equals the measured value and Y equals the value after adjustment.