beautypg.com

Sonics VC750 (Serial No. "Y through "AB")" User Manual

Page 6

background image

Displays prompts and the following control parameters:

• Amplitude selected

• Output power delivered to the probe in watts, and as percentage of the total

power

• Selected duration of processing

• Actual processing time

• Elapsed time

• Set and read temperature

• Pulse duration

• Accumulated amount of energy in Joules delivered to the probe.

Input digits.

Clears preceding entry.

Enters data into the program, and selects various parameters, for display on the
LCD screen.

Used with the numeric keys to set the duration of ultrasonic application – from 1
second to 9 hours, 59 minutes, 59 seconds.

Used with the numeric keys to set the pulse mode. The ON cycle and OFF cycle
can be set independently from 1 second to 59 seconds. Red indicator lights when
pulser is in the OFF portion of the cycle.

Starts or stops the ultrasonics. In the STOP mode the red indicator goes off.

Suspends operation. Red indicator lights when the processing cycle is interrupted.

Controls the amplitude of vibration at the probe tip.

Switches the main power on.

Switches the main power off.

Used with the AMPL key when the unit is on stand-by to set the amplitude of
vibration at the probe tip. Also used to increase or decrease the amplitude in small
increments while the unit is running. To accomplish this task, depress the AMPL
key to display the percentage of amplitude previously selected, then depress the

or key as required.

▲ ▼

key

I

key

AMPL key

PAUSE key

START/STOP key

PULSER

key

ENTER/REVIEW

key

0 – 9 key

TIMER

key

CLEAR key

FUNCTIONS OF KEYS, CONTROLS, INDICATORS, AND CONNECTORS

FRONT PANEL

3

Screen

O key

This manual is related to the following products: