PREPARATION FOR USE

1. Plug the electrical line cord into the electrical outlet. If the unit is already on, switch

the unit off by depressing the O key.

2. If the optional footswitch is used, insert the footswitch plug into the jack located

on the rear panel. Make sure that the plug is inserted all the way in.

3. Screw the booster into the converter and secure forcefully with the spanner

wrenches provided.

4. Ensure that the High Volume Flow Cell is completely drained out of any liquid.

5. Using the spanner wrenches provided, forcefully secure the booster to the High

Volume Flow Cell probe.

6. Connect the converter cable to the power supply.

7. Connect either one of the air fittings on top of the converter to a source of dry

compressed air.

CAUTION

Never place a washer between the converter,

booster or probe. Never apply grease to the

mating surfaces or threads of the converter,

booster or probe.

CAUTION

Do not operate an Ultrasonic Processor that has

been in a very cold or hot environment for a

prolonged period of time. Wait until it has

reached room temperature

NOTE

The High Volume Flow Cell is recommended only for the treatment of

low viscosity samples, which do not require extended exposure to

ultrasonics. Because the residence time within the cell is relatively

short, it might be necessary to recirculate the sample.