Sonics VCX1500HV User Manual
Page 13
PULSER:* The pulse function is not used when working with the High Volume Flow
Cell, but can be used when processing a batch with a probe. By inhibiting heat build-
up in the sample, the pulse function enables safe treatment of temperature sensitive
samples at high intensity. In addition, pulsing enhances processing by allowing the
material to settle back under the probe after each burst. The ON and OFF pulse
duration can be set independently from 1 second to 59 seconds. During the OFF
portion of the cycle, the red indicator on the PULSER key will illuminate. If the OFF
portion of the cycle exceeds three seconds, a cautionary message – CAUTION –
PROBE ON STANDBY – will warn the operator against touching the ultrasonic probe.
To set the pulser, press the PULSER key.
The screen will display:
Using the numeric keys set the ON portion of the cycle, then press the
ENTER/REVIEW key.
The screen will display: e.g.
Using the numeric keys set the OFF portion of the cycle.
The screen will display: e.g.
Press the ENTER/REVIEW key.
The screen will display:
* Should not be used when working with the High Volume Flow Cell.
10
Pulse on __.__sec
Pulse off __.__sec
Pulse on 01 sec
Pulse off __.__sec
TIME 5:30:25
TEMP __ __°C
PULSE 01 01
AMPL
40%
On Cycle
Off Cycle
Pulse on 01 sec
Pulse off 01 sec