beautypg.com

Sonics VCX750 User Manual

Page 14

background image

9

Press the I key. The screen will display the power rating and the frequency of the
Ultrasonic Processor, and the following control parameters.

AMPLITUDE: The amplitude is the only parameter that must be set in order for the
Ultrasonic Processor to be operational. The other control parameters – Time and
Pulse, do not have to be set for continuous operation. AMPL. displays the percentage
of amplitude that was previously selected. Press the AMPL key and the

or

key

for a reading of 40%, then depress the ENTER/REVIEW key.

The screen will display:

The Ultrasonic Processor is now ready for continuous operation. To energize the
equipment, press the START key or the footswitch. To de-energize the equipment,
press the STOP key or release the footswitch. If the Time or Pulse* functions must be
used, refer to the appropriate paragraph(s) below.

To increase or decrease the amplitude in small increments when the equipment is on,
depress the AMPL key to display the percentage of amplitude that was previously
selected, then depress the

or

key, as required.

CAUTION

Do not operate the power supply unless it is

connected to the converter.

TIME __:__:__

TEMP __ __°C

PULSE _ _ _ _

AMPL __ __%

NOTE

Any combination of functions can be selected in

any order. To clear an erroneous entry press

the CLEAR key.

TIME __:__:__

TEMP __ __°C

PULSE _ _ _ _

AMPL

40%

This manual is related to the following products: