Set the over temperature limit control, Starting the auto cycle, Ending the auto cycle – Shellab SRI20PF User Manual
Page 23: Operation (continued)

23 |
P a g e
OPERATION (CONTINUED)
S
ET THE
O
VER
T
EMPERATURE
L
IMIT
C
ONTROL
Perform the following steps to set up the Over Temperature Limit system for use after calibrating
the unit.
1. If you have not done so already, turn the Set Over Temperature Limit control dial
clockwise to the maximum position. This allows the Set Temperature control to stabilize.
2. Then turn the Over Temperature Limit control dial counterclockwise until the Over Temp
Limit Activated light illuminates.
3. Slowly turn the Over Temperature Limit control dial clockwise until the light turns off.
4. This adjusts the Over Temperature Limit control to approximately 1
˚C above the
temperature configured by the Set Temperature control.
Note: Test the Over Temperature feature annually
S
TARTING THE
A
UTO
C
YCLE
To start the incubator’s day – night Auto Cycle Mode carry out the following step:
1. Press and hold the Cycle Start button for two seconds.
2. The Program Active LED indicator next to the Cycle Start button should illuminate and stay
lit.
E
NDING THE
A
UTO
C
YCLE
To end the Auto Cycle perform the following step:
1. Push and hold the Cycle Start button for two seconds
2. The Program Active indicator next to the Cycle Start button should extinguish.
Note: In the event of a power outage, the unit will restart in the last mode it was operating in, with
the Current Time display automatically adjusted to compensate. If running in the Auto Cycle
mode, the incubator will advance itself to the correct Day Time or Night Time phase of the
automatic temperature and illumination cycle.