Mechanical – Shellab SCO10A-2 User Manual

Page 16

background image

16

Overshoots set point and continues to rise
- display and Fyrite match

Debris in solenoid causing it to leak continuously

Rises very slowly

1/ Filter overly dirty or partially plugged

2/ Hose kinked or leaking

3/ CO

2

tank regulator set too low.

4/ CO

2

tank near empty

Never rises

1/ CO

2

filter plugged

2/ CO

2

hose blockage

3/ CO

2

tank empty

4/ Set point is at 0.0 and has not been reset.

5/ CO

2

tank regulator not on

Display and Fyrite reading do not match

1/ Old Fyrite fluid.

2/ Out of calibration.

Unstable display

1/ Confirm that fan is working

2/ Magnetic field interference

3/ Electronic problem with CO

2

sensor, interface or controller

4/ Top of unit exposed to cold air drafts

5/ Incubator too heavily loaded

Feeding continuously or abnormally high
CO

2

usage

1/ Do decay test: if more than 1% decay in an hour, check for leak : door

gasket tightness, motor inlet to chamber, sensor and probe inlet to
chamber

2/ Door being opened too often

Won’t hold calibration – Fyrite reading
varies but display stable

1/ Atmospheric pressure fluctuations

2/ Top of unit exposed to cold air drafts

3/ Unit being operated without shadow box cover in place

4/ Condensation collecting on CO

2

sensor

5/ CO

2

sensor or interface failure

6/ Unit incorrectly calibrated.

7/ Taking Fyrite reading too soon after the door has been opened

8/ Air leak around CO

2

sensor mounting plate

MECHANICAL

Door not sealing

1/ check physical condition of gasket.

2/ Confirm that door latch pulls door in tightly.

3/ Assure that gasket is in original location.

Motor doesn't move

1/ if shaft spins freely: check connections to motor and check voltage to

motor;

2/ if shaft rubs or is frozen, relieve binding and retest

Motor makes noise