HEPA CO
2
FILTER
The CO
2
Incubator design features a HEPA filtration system with a patented copper
housing. (Patent Number 6,333,004) As the chamber air is drawn through the filter 
system, airborne microbes and isolated particulates are not only removed from the air, 
but destroyed as well. It is through copper oxidation that the contaminants are actually 
killed. The HEPA filter is easily replaced without the need of any tools and has an 
efficiency rating of 99.97% at 0.3 microns. 
Install the HEPA Filter onto the duct prior to installing the duct in the unit.
WARNING: Ensure that when installing the duct to ensure that the duct is not 
touching the C02 sensor head as this may cause erroneous readings.
2800517
.
Section
8
The Hepa filter attaches to a fixed 
coupling ring with a direct “nap-in” 
action. Push straight up to the 
coupling ring to connect. Pull the 
filter straight down to disengage. 
Replacement 
Hepa
filters
is
available from Sheldon Mfg., Inc. 
Part Number 2800517. 