Daily maintenance, Normal gas consumption, User maintenance (continued) – Shellab BACTRON600 User Manual
Page 51

51 |
P a g e
USER MAINTENANCE (CONTINUED)
D
AILY
M
AINTENANCE
1. Exchange the catalyst cartridge with a reactivated one.
2. Remove and empty the condensate collection container as needed.
3. Change the anaerobic indicator strip at least once per 24 hours, or as often as necessary
so that the indicating strip in the workspace chamber remains moist.
4. Visually inspect that the pass box gaskets are properly seated.
5. Check the cuffs on the sleeve system for holes, tears, and other signs of wear that may
compromise integrity. Replace if necessary.
6. Check the incubator temperature to ensure that the setting is correct.
7. Record the gas cylinder reading(s).
8. Change or reactivate the charcoal scrubber, if installed. Please see the
Charcoal Scrubber and Rejuvenation Cycle
on page 48 in the Operation section.
9. Clean and disinfect the workspace chamber after each use, or in accordance with your
laboratory or production protocols, or regulatory requirements.
N
ORMAL
G
AS
C
ONSUMPTION
A sealed and undisturbed BACTRON workstation will typically go for more than 30 minutes
between gas injections into the workspace chamber. When injecting, the Chamber Gas pilot light
will illuminate, accompanied by a pair of audible clicks from the activated gas solenoid.
The manometer water column should be depressed by approximately half an inch (1 cm) under
normal operating conditions.
Injections taking place every thirty minutes or less in an undisturbed unit may indicate a small leak.
Injections every 10
–
30 seconds are indicative of a large leak.
Note: Pass box cycles, entering or exiting the arm port doors, or working within the workspace
chamber may temporarily increase the frequency of gas injections.