Control panel overview (continued) – Shellab BACTRON600 User Manual
Page 23

23 |
P a g e
CONTROL PANEL OVERVIEW (CONTINUED)
Workspace Incubator Control Panel (BACTRON900 / BACTRONIV-900 Only)
Figure 5: BACTRON900 Workspace Control Panel
The workspace incubator control panel on the BACTRON900 controls the operations of the
workspace
chamber incubator and the incubator’s Over Temperature Limit system.
Main Temperature Control and Green Digital Display
The BACTRON900’s
workspace SET TEMPERATURE control panel comes with a green digital
display that shows the temperature of the incubator accurate to within + 0.1
C. The control consists
of up and down arrow keys for adjusting set point temperatures or placing the display in its
calibration mode.
Heating Activated Light
The clear pilot light located beneath the label HEATING ACTIVATED illuminates whenever the
heating elements are powered and warming the BACTRON900’s
workspace chamber Incubator
Set Over Temperature
This graduated dial sets the temperature set point for the Over Temperature Limit backup system in
the workspace chamber of the BACTRON900. The OTL System operates independently of the
Main Temperature Controller, and prevents uncontrolled heating of the workspace incubator in the
event of a temperature control board failure. For more details, please see the explanation of
OTL Light
This pilot light is marked OVERTEMPERATURE ACTIVATED and is located directly below the
Over Temperature Limit control. The light will turn on and a buzzer alarm will activate when the
OTL System has taken control of the incubator. Under normal operating conditions this light should
not illuminate.