Environmental requirements, Specifications sa series analytical balances – Scientech S9000 User Manual
Page 21

ENVIRONMENTAL REQUIREMENTS
This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 1 or 2 in accordance with IEC 664 and transient overvoltages according to Installation
Categories (Overvoltage Categories) I, II and III. For mains supply, the minimum and normal category is II.
SPECIFICATIONS SA SERIES ANALYTICAL BALANCES
Model
SA 80
SA 120
SA 210
SA 310
SA 410
SA 510
SA 210D
SA 525D
Weighing Mode
Single Range
Dual Range
(Floating
Tare™)
Capacity
80g
120g
210g
310g
410g
510g
100g 250g
200g
500g
Readability
.0001g
.0001g
.001g
Tare Range
0-80g
0-120g
0-210g
0-310g
0-410g
0-510g
0-200g 0-500g
0-200g
0-500g
Repeatability
.0001g
.0001g
.0001g
.00015g
.00015g
.0002g
.0001g
.00015g
(Standard
Deviation)
.0005g
.0005g
Linearity
±.0002g
±.0002g
±.0005g
Stabilization Time
User Selectable
Weighing Speed
Adaptive
Display Update
Interval
200ms
Data Interface
RS-232 bidirectional interface is standard. Baud rates are 300, 600, 1200, 2400, 4800, 9600 and
19,200.
Pan Size
Circular 3.5 inch diameter
Operating
Temperature
5°C to 40°C
Relative Humidity
80 % for temperatures up to 31°C decreasing linearly to 50 % at 40°C
Power Supply
115 Volts/60 Hz ± 10 %
230 Volts/50Hz or 60 Hz ± 10 %
Windscreen
7.5W x 8.25D x 9.5H
Calibration
Weight
80g
100g
200g
300g 400g
500g
200g
500g
50g
50g
100g
200g
200g
100g
200g
Housing (WxDxH)
7.5" x 11.25" x 3.25"
Weight
~ 13.8 lbs
21