Scientech S9000 User Manual
Page 13

13
4. LOCAL CONTROL SELECTION (LC on is
default). “LC on” means that the front panel
switches control the balance. To enable a
computer to control the balance, “LC oFF”
must be selected.
Press the MODE switch.
OK and the current front panel control setting , LC on or LC oFF is
displayed.
Press the SEND switch to choose between
on or off. Then press the ZERO switch.
OK SAVEd is displayed followed by OK and the chosen setting.
5. PROTOCOL or HANDSHAKE SELECTION
(Pr on is default).
Press the MODE switch.
OK and the current software protocol setting, Pr oFF or Pr on is
displayed.
Press the SEND switch to choose between
on or off. Then press the ZERO switch.
OK SAVEd is displayed followed by OK and the chosen setting.
6.
PARITY SELECTION (PA nonE is default).
Press the MODE switch.
OK and the current parity setting, PA EVEn, PA odd, or PA nonE is
displayed.
Press the SEND switch to choose between
EVEn, odd, or nonE. Then press the
ZERO switch.
OK SAVEd is displayed followed by OK and the chosen setting.
7. AUTO SEND SELECTION (AS oFF is
default). “AS on” means the balance will
automatically send the display contents over
the RS-232 interface at the selected time
interval placed into register 89. This time
interval may be inputted via the optional PDA
or through Type 4 RS-232 commands. Refer to
these sections for instructions. A time and date
stamp may also be sent at the same time (see
Type 4 RS-232 Commands section). The auto
send selection is not functional if the send-
when-stable command is on (see Selectable
Filter section).
Press the MODE switch.
OK and the current auto send setting, AS oFF or AS on is displayed.
Press the SEND switch to choose between
on or off. The press the ZERO switch.
OK SAVEd is displayed followed by OK and the chosen setting.
8. CONTINUOUS SEND SELECTION (CS oFF
is default). “CS on” means the balance will
automatically send the weight reading over the
RS-232 interface at the baud rate selected.
Display readings are updated every 200
milliseconds. The continuous send selection is
not functional if the send-when-stable
command is on (see Selectable Filter section).
Press the MODE switch.
OK and the current continuous send setting, CS oFF or CS on is
displayed.
Press the SEND switch to choose between
on or off. Then press the ZERO switch.
OK SAVEd is displayed followed by OK and the chosen setting.
9.
DISPLAY BLANKING SELECTION (bL on is
default). “bL on” means that every time data is
sent over the RS-232 interface, the display will
blink.
Press the MODE switch.
OK and the current display blanking setting, bL on or bL oFF is
displayed.
Press the SEND switch to choose between
on or off. Then press the ZERO switch.
OK SAVEd is displayed followed by OK and the chosen setting.
10. Press the ON/OFF switch.
The display blanks
11. Press the ON/OFF switch.
The balance returns to normal operation with the new RS-232
configuration.