4. LOCAL CONTROL SELECTION (LC on is

default). “LC on” means that the front panel

switches control the balance. To enable a

computer to control the balance, “LC oFF”

must be selected.

Press the MODE switch.

OK and the current front panel control setting , LC on or LC oFF is

displayed.

Press the SEND switch to choose between

on or off. Then press the ZERO switch.

OK SAVEd is displayed followed by OK and the chosen setting.

5. PROTOCOL or HANDSHAKE SELECTION

(Pr on is default).

Press the MODE switch.

OK and the current software protocol setting, Pr oFF or Pr on is

displayed.

Press the SEND switch to choose between

on or off. Then press the ZERO switch.

OK SAVEd is displayed followed by OK and the chosen setting.

6.

PARITY SELECTION (PA nonE is default).

Press the MODE switch.

OK and the current parity setting, PA EVEn, PA odd, or PA nonE is

displayed.

Press the SEND switch to choose between

EVEn, odd, or nonE. Then press the

ZERO switch.

OK SAVEd is displayed followed by OK and the chosen setting.

7. AUTO SEND SELECTION (AS oFF is

default). “AS on” means the balance will

automatically send the display contents over

the RS-232 interface at the selected time

interval placed into register 89. This time

interval may be inputted via the optional PDA

or through Type 4 RS-232 commands. Refer to

these sections for instructions. A time and date

stamp may also be sent at the same time (see

Type 4 RS-232 Commands section). The auto

send selection is not functional if the send-

when-stable command is on (see Selectable

Filter section).

Press the MODE switch.

OK and the current auto send setting, AS oFF or AS on is displayed.

Press the SEND switch to choose between

on or off. The press the ZERO switch.

OK SAVEd is displayed followed by OK and the chosen setting.

8. CONTINUOUS SEND SELECTION (CS oFF

is default). “CS on” means the balance will

automatically send the weight reading over the

RS-232 interface at the baud rate selected.

Display readings are updated every 200

milliseconds. The continuous send selection is

not functional if the send-when-stable

command is on (see Selectable Filter section).

Press the MODE switch.

OK and the current continuous send setting, CS oFF or CS on is

displayed.

Press the SEND switch to choose between

on or off. Then press the ZERO switch.

OK SAVEd is displayed followed by OK and the chosen setting.

9.

DISPLAY BLANKING SELECTION (bL on is

default). “bL on” means that every time data is

sent over the RS-232 interface, the display will

blink.

Press the MODE switch.

OK and the current display blanking setting, bL on or bL oFF is

displayed.

Press the SEND switch to choose between

on or off. Then press the ZERO switch.

OK SAVEd is displayed followed by OK and the chosen setting.

10. Press the ON/OFF switch.

The display blanks

11. Press the ON/OFF switch.

The balance returns to normal operation with the new RS-232

configuration.