Scientech S9000 User Manual
Page 15

15
[decimal number]_CONVERT[CR]
None
Loads the number entered, as a "last
front panel units" relative multiplier,
into the CONVERT register and enters
the CONVERT mode. CLEAR exits the
convert mode and undefines the convert
register.
[decimal number]_CAL[CR]
None
Loads a computed number based on the
entered number, as a "last front panel
units" relative multiplier, into the CAL
register and enters the CAL mode. The
display must be non-zero when this
command is issued. If the display is
zero, a 3 second "Err" message is
displayed and the balance returns to its
normal weighing mode. The current
weight reads as the number entered with
the appropriate sign. CLEAR exits the
CAL mode and undefines the CAL
register.
[integer]_PIECES[CR]
None
Loads a computed number based on the
entered number, as a "last front panel
units" relative multiplier, into the
PIECES register and enters the PIECES
mode. The display must be non-zero
when this command is issued. If the
display is zero, a 3 second "Err"
message is displayed and the balance
returns to its normal weighing mode.
CLEAR exits the pieces mode and
undefines the pieces register.
[Decimal
number]_ENTER_[integer]
_PIECES[CR]
None
Accepts the decimal number as a weight
in "last front panel units" per integer
pieces. If the balance is not in PCS
mode, it goes into PCS mode using the
decimal number/the integer as the
weight. If used when the balance is
already in PCS mode, the display digits
are blanked for about 1 second while the
PCS an "last front panel units"
annunciators are lit. The balance then
resumes normal PCS operation using the
decimal number/the integer as the unit
weight.
[register number]_RCL_[(CONVERT)
(CAL) (PIECES)][CR]
REG XXX :[register contents][MULT.]
Puts the recalled register contents into
the specified mode register and enters
that mode.
TAVG[CR]
None
Clears the TBAR registers and enters
the TBAR mode.
RESET_TAVG[CR]
None
Clears the TBAR registers. If the
balance is in the TBAR mode, the
process keeps running.
XAVG[CR]
None
Accumulates the current weight in the
XBAR registers.
RESET_XAVG[CR]
None
Clears the XBAR registers.
GLP[CR]
Sends GLP header
Puts balance in the GLP mode.
XAVG[CR]
Sends weight while in GLP
Accumulates the current weight in the
XBAR registers.
RCL_XAVG[CR]
Sends statistics and signature line
Terminates GLP mode.
ISO[CR]
Sends ISO header and waits for target
weight entry
Puts balance in ISO mode.
xxx.xx_Targ[CR]
Sends target weight (g)
Enter target weight value in grams.