KERN & SOHN MWS 400K100DM User Manual
Page 57

Position of adjustment switch:
Position of adjustment switch Status
1. left side
not documented
2. centre
adjustment position - adjustment is
possible
3. right side
verification position – adjustment locking
13.3 Checking the scale settings concerning scale verification
To start the adjustment function, a scale is to be switched over into the service
mode. To do this, set the adjustment switch in the adjustment position (refer to
chapter 12.2).
The service mode makes it possible to change all parameters of a scale. Service
parameters cannot be changed because it may have an influence on the scale
settings.
13.3.1 Menu overview in the service mode (adjustment switch in the adjustment
position)
Overview is only used to check the set parameters by authorised verification offices.
Changes may only be introduced in parameters of the automatic switching off
function „
A.OFF” and audible signal „BURR”.
1 2 3
MPS / MTS / MXS / MWS-BA-e-1417
57