6.2 Electromagnetic interferences
Guidelines and manufacturer’s declaration – electromagnetic interferences
The baby scale MBC-M is designed for use in an electromagnetic environment that meets the
requirements stated below. The customer or user of the medical electrical device must ensure that
operation takes place in such an environment.
Emitted interference
measurements
Conformity
Electromagnetic environment -
guideline
HF emissions
as per CISPR 11 / EN 55011
Group 1
The baby scale MBC-M uses HF
energy merely for its internal working.
Its HF emission therefore is very low
and it is highly unlike to interfere with
adjacent electronic devices.
HF emissions
as per CISPR 11 / EN 55011
Class B
The baby scale MBC-M is designed
for use in all equipment including
those in living areas and those
connected directly to the public power
grid that also supplies buildings used
for living purposes.
Emission of harmonics
as per IEC 61000-3-2
Class A
Emission of voltage fluctuations /
flicker
as per IEC 61000-3-3
Conforms with
Do not put the baby scale MBC-M directly next to other devices and do not stack it
with other devices. If this type of operation is necessary, observe the baby scale
MBC-M to ensure normal operation in such an arrangement.